I. Scope of Application
1.1 It is applicable to the test for evaluating the surface
combustion performance of materials by radiant heat flux.
1.2 ZY6125-PC, also known as radiant plate heat flux tester, is one
of the most widely used methods for determining flame propagation
performance in laboratories. The specimen is placed at a 30° Angle
to the radiation plate, 120mm from the upper edge and 340mm from
the lower edge of the radiation plate. The ignition source is a
medium-sized gas blowtorch. The blowtorch is a 230mm long ceramic
tube with a diameter of 6mm, placed at an Angle of 15° to 20° to
the test piece. During the test, the specimen is exposed to the
heat source of the radiation plate and a medium-sized blowtorch for
up to 15 minutes. After the specimen is ignited, the time it takes
for the flame front to reach the reference mark is recorded. In the
experiment, the amount of heat radiation released by the flue gas,
the smoke concentration, the droplets from combustion and the flame
spread index Is were recorded. Calculate the radiation plate index
Is equal to the factor Fs of sample flame spread multiplied by the
factor Q of heat change, that Is, Is=Fs * Q
1.3 This testing machine can also be used to evaluate the flame
spread performance of components such as interior panels, ceilings,
sound insulation materials, Windows, door lintels, and seat side
panels of trains in accordance with the NFPA 130 standard for flame
retardancy of rail vehicles in the United States. Judgment
criteria: The test duration should be no more than 15 minutes, or
the flame should reach 380mm at the reference mark.
Ii. Compliance with Standards
2.1 Comply with ASTM E 162-08B "Standard Test Method for Surface
Flammability of Material Using a Radiant Heat Energy Test Method
for Evaluating the surface burning performance of materials using
radiant heat sources.
Iii. Main Parameters
3.1 Radiation Section:
3.1.1 Radiant panel: Composed of porous refractory materials, the
exposed size of the radiant panel is 305mm×457mm, and it can
withstand temperatures above 815℃. 3.1.2 Air compressor: Generates
an air flow rate of 3000L/min; The air pressure is 2.8 inches of
water column (700Pa).
3.1.3 Gas Supply pipeline: Install air filters, pressure regulators
and stop valves to regulate the gas flow
3.1.4 Ignition System: High-voltage electronic ignition;
3.2 Sample Holder:
3.2.1 Sample holder: Made of heat-resistant chromium steel, with a
3-inch (76mm) interval line observation mark engraved on the
surface of the sample holder. 3.2.2 Support frame for the sample
holder: This frame will have two stainless steel crossbars, each
with a diameter of 0.5±0.13 inches (12.7±3.3mm), allowing the
sample to be directly fixed in front of the radiation plate. The
supports and support components are made of metal. Since the
installation Angle between the sample and the radiation plate is of
crucial importance, the specified dimensions of the frame are made
strictly in accordance with the dimensions marked in ASTM E162-08b
as shown in Figure 1, and the tolerance is controlled within 0.125
(3.2mm).
3.3 Test blowtorch: Outer diameter ¢4.8mm, inner diameter ¢3.2mm,
length 205mm. To extend the service life of the test blowtorch, a
porcelain tube sleeve was installed on the part of the blowtorch
exposed to the radiation source. The inner diameter of the
porcelain tube is 5.2mm and the outer diameter is 7.14mm. The
blowtorch is installed horizontally and placed at an Angle of 15°
to 20° to the horizontal plane of the sample. When not in use, the
blowtorch can be removed. The blowtorch is composed of a Venturi
mixer, where acetylene and air are pre-mixed in the Venturi. The
flame height of the combustion burner is 76mm, and the distance
between the burner and the central surface of the upper part of the
sample is 12.7mm.
3.4 Chimney: Made of 0.040-inch (1.0mm) stainless steel plate, the
shape and size are as shown in Figure 1. The installation position
of the chimney with the sample and the radiation plate is strictly
designed in accordance with Figure 1.
3.5 Thermocouples: Eight thermocouples are installed at equal
intervals and in parallel on the chimney. The installation
positions are shown in Figure 1. K-type stainless steel armored
thermocouple with a diameter of 3.2mm; The temperature-sensing wire
is 0.5mm and can measure a maximum temperature of 1200°C. The
verification of the rated unit heat output of the blowtorch is
carried out by regularly calibrating the temperature of the
thermocouple on the chimney, and the verification procedure is
executed in accordance with A1.2.
3.6 Data Acquisition System:
3.6.1 Collect and record the real-time temperature curves and data
of the thermocouple on the chimney from 38°C to 538°C, with a
collection frequency of once every 5 seconds.
3.6.2 Computer data acquisition system: The accuracy of temperature
acquisition is 0.01%
3.6.3 The frequency of all data collection, recording and storage
shall be once every 5 seconds and continuously for one hour.
3.7 Exhaust hood: Install an exhaust hood with a fan on the smoke
and dust. When the radiant panel is not in operation, the fan at
the top of the chimney can produce a wind speed of 100 feet per
second (0.5M/S) (30.5M) per minute. When the radiant panel is in
operation, the fan can generate approximately 250 feet per minute
(78M/min).
3.8 Radiation Pyrometer:
3.8.1 Adopt high-precision radiation pyrometers from Swiss KELLER
Company;
3.8.2 Measuring range: (480-530) ℃ blackbody temperature;
3.8.3 Measurement Accuracy: ±0.3℃
3.8.4 Sensitivity: Constant within the wavelength range of 1 μ m to
9 μ m;
3.8.5 Installation position: Approximately 1.2 meters away from the
radiant panel, capable of sensing temperature to a circular surface
with a diameter of 254mm above the radiation.
3.8.6 The calibration of the blackbody temperature range and
procedures of the radiation pyrometer shall be carried out in
accordance with Appendix A1.
3.9 Timer: Resolution 0.01min, accuracy 1 second /h.
3.10 Heat flux meter
3.10.1 Measuring range: (0-15) Kw/m ²
3.10.2 Accuracy of heat flux meter: ±0.2Kw/m ²;
3.10.3 Heat flux meter accuracy: <±3%;
Iv. Equipment Structure
4.1 The structure of this machine consists of two parts: the
combustion box and the control box.
4.2 Material of control box/combustion box: Made of high-quality
stainless steel plate, processed by CNC machine tools, with an
elegant and generous arc shape.
4.3 Other mechanical parts are made of high-quality stainless steel
or A3 material with thickened electroplating to prevent corrosion
and rust.
V. Control System:
5.1 Control Mode: The board card module I/O board is adopted, with
PID+SSR control mode. The acquisition system can collect and record
the CHF value of the radiation flux curve, as well as the flame
extinction time and the distance of flame propagation.
5.2 Status Value display: The software automatically has the status
recognition function for conducting tests, such as when the heat
flow rate meets the standard requirements; 5.3 Heat Flow meter
calibration procedure, step-by-step calibration;
5.4 Printing of record, test and calibration reports; Record and
print data every 5 seconds (at least within 1 hour);
5.5 Computer: One laptop
|